22
Installation
Water cooling
The turbopumps HiPace 800 P with TC 400 have water cooling as standard equipment.
●
In case of increased backing pressure (> 0.1 hPa) and/or operation with high gas
throughputs, either air or water cooling may be used.
●
Generally use water cooling if the ambient temperature is > +35 °C.
Cooling water requirements
Connecting to a cooling water system
Insert hoses for the water cooling fore- and return-line as far as they will go in one
cooling water connection each.
Recommendation:
install dirt trap in the fore-line.
Cooling water connection
Socket connection
Hose lines
External diameter 8 mm
Internal diameter 6 mm
Cooling water quality
filtrated, mechanically clean, optically
clear, no turbidity, no sediments, chemical-
ly neutral
Oxygen content max.
4 mg/kg
Chloride content max.
100 mg/kg
Water hardness max.
10 °dH
12.53 °e
17.8 °fH
178 ppm CaC0
3
Consumption of potassium permanganate max.
10 mg/kg
Carbon dioxide content max.
undetectable
Ammonia content max.
undetectable
pH-value
7 - 9
Fore-line overpressure max.
600 kPa
Cooling water temperature
refer to "Technical Data"
Cooling water consumption at max. gas throughput
refer to "Technical Data"
Fig. 11: Connecting the cooling water
14
Cooling water connection
14a Fore-line
14b Return line
14c Dirt trap
14
14a
14b
14c