6. Pull the filter unit completely out of the operating fluid separator.
7. Dismantle the clamping plate.
8. Replace the exhaust filters.
7.4.2 Mounting the exhaust filters
Prerequisites
● Vacuum pump switched off
● Vacuum pump vented to atmospheric pressure
● Vacuum pump cooled
Required consumables
● Maintenance kit – Maintenance level 2
Required tools
● Socket key,
WAF 13
● Allen key,
WAF 6
● Calibrated torque wrench (tightening factor ≤ 2.5)
8x
2
1
1.1
3.1
3
4
5
6
8
7
9
Fig. 19:
Mounting the exhaust filters
1
Exhaust filter
5
Filter material
1.1 O-ring
6
Fixing bolts
2
Clamping plate
7
Flat seal
3
Filter unit
8
Separator cover plate
3.1 O-ring
9
Screws
4
Operating fluid separator
Procedure
1. Place the new O-ring on the front side of the exhaust filters.
2. Mount the exhaust filters using the clamping plate.
– The installation orientation is identifiable by an arrow on the filter.
– The arrow must point upwards (
↑
).
3. Place the new O-ring on the front side of the filter unit.
4. Insert the pre-assembled filter unit into the operating fluid separator.
5. Insert the new filter material into the housing groove of the operating fluid separator and fix it with
the fixing bolts at the same time.