5.6.1 Solenoid valve for gas ballast valve (standard version)
7
6
2
1
3
4
5
10
8
9
Fig. 15:
Solenoid valve connection for gas ballast valve
1
Fixing screw
6
Screws
2
Gas ballast knob
7
Dosing screw
3
Screws
8
O-ring
4
Flange
9
O-ring
5
Conversion kit
10
Protective cap
Required tools
● Screwdriver
● Allen key, 4 mm
● Torque wrench
Remove gas ballast valve from pump
1. Unscrew fixing screw from gas ballast knob.
2. Remove gas ballast knob.
3. Unscrew both screws from flange.
4. Pull the valve flange axially out of the housing.
Installing conversion kit with solenoid valve
1. Insert solenoid valve axially into the housing.
2. Be careful with the O-rings.
3. Tighten the screws,
– torque:
3 Nm
.
4. Open dosing screw by rotating to the left up to the stop.
5. Remove protective cap.
6. Establish the electrical connection.
7. Install the protective cap.
Order number
PK 223 717 -U
Classification advanced
Gas ballast valve conversion kit with magnetic valve
Flange (in)
G 1/4
Over pressure max.
500 hPa
Electrical connection
Cable screw connection PG9
For pump
Duo 35/65
Power supply voltage
24 V DC
Contact
Normally closed
Installation
26/66