44
Troubleshooting
Troubleshooting R-net
The following troubleshooting guide describes a number of faults and events which may occur when you
use your wheelchair, together with suggested remedies. Note that this guide cannot describe all the
problems and events which may occur and you should always contact your service contact or Permobil
in case of doubt.
The wheelchair will not start.
Batteries discharged.
Charge the batteries.
The cable connection to
Insert the cable in the control
the control panel has
panel.
come loose.
Main fuse blown
.
Check possible causes carefully
before resetting/replacing the
main fuse. See page 40.
The wheelchair cannot be
Battery charger connected.
Stop charging and disconnect
driven.
the charging cable from the
wheelchair’s charging socket.
Brake release activated.
Reset the brake release.
The wheelchair is locked.
Unlock the wheelchair.
See user manual
An exclamation mark on the
Electronics fault.
See pages 45–58.
control panel display is
flashing rapidly and the
wheelchair will not run.
The wheelchair can only be
Seat lift raised too high.
Lower seat lift.
driven at reduced speed.
The wheelchair cannot be
Main fuse blown
.
Check possible causes carefully
charged.
before resetting/replacing the
main fuse. See page 40.
The wheelchair “switches
The electronics’ energy-saving Switch the wheelchair on again
itself off” after a certain period mode has been activated.
using the start key on the control
of inactivity.
panel.
EVENT
POSSIBLE CAUSE
REMEDY
Содержание C350
Страница 1: ...Permobil C350 SERVICE MANUAL Power Wheelchair US ...
Страница 2: ......
Страница 62: ...62 Cabling overview A B C D E F 1 2 3 4 ...
Страница 63: ...63 Cabling overview Cabling overview R net A B C D E F 5 6 7 8 ...
Страница 64: ...64 Cabling overview A B C D E F 1 2 3 4 ...
Страница 65: ...65 Cabling overview Cabling overview VR2 A B C D E F 5 6 7 8 ...
Страница 67: ...67 ...
Страница 68: ...Article no 205219 US 0 ...