22
O
2
asis User Guide
PERFORMANCE
Startup time to normal use: 30 seconds
Tests automatically performed at startup: Temperature check, sensor check,
consumable present check, last setting check.
Passive humidification begins immediately. Full heated humidification perfor-
mance can take up to 30 minutes at start-up, depending on flow rate.
Units of Measure:
Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ml (milliliters)
Flow Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . lpm (liters per minute)
Leakages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ml (milliliters)
Oxygen Supply Flow Rate Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.5 lpm to 6 lpm
Maximum Oxygen supply flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 lpm
ESTIMATED HUMIDIFICATION OUTPUT
Dry Oxygen
Flow Rate
Heat Setting
20°C Ambient Temp
35°C Ambient Temp
Outlet Dew
Point (°C)
Outlet
Humidity
(mg/l) BTPS
Outlet Dew
Point (°C)
Outlet
Humidity
(mg/l) BTPS
0.5
Unplugged
15
11
26
23
Low
29
27
36
40
Medium
30
28
37
42
High
32
32
40
49
3
Unplugged
12
9
22
18
Low
26
23
25
21
Medium
30
28
30
28
High
34
36
34
36
6
Unplugged
8
7
17
13
Low
18
14
19
15
Medium
22
18
23
19
High
24
20
25
21
PP-19-1 REV. 15