asis User Guide

Problem

Possible Cause

Action to Take

The device does 
nothing after 
providing power. 
No LED light in-
dicators are on.

The device is 
unplugged, or the 
outlet is not pro-
viding power.

Check that the AC power adapter is properly 
plugged in to an electric outlet and that the 
outlet has power available to it. Also check that 
the barrel connector at the end of the adapter 
cord is securely connected to the power input 
jack on the base unit. If the device still does 
not power on, or power is intermittent, please 
contact the care provider, licensed distributor, 
or O2asis to return the device.

The air flow 
coming out of 
the cannula feels 
too low.

The connection 
between the 
oxygen input hose 
and the device is 
not secure, or the 
connection be-
tween the cannula 
and the heated 
line is not secure.

Check that the oxygen input connector is 
securely attached to the device’s oxygen 
input port. Push on until the connector is snug 
with the device. Also check that the cannula 
connector is secured to the heated line. Push 
the end of the cannula into the open end of the 
heated line until snug. If the device still does not 
provide the proper air flow, please contact the 
care provider, licensed distributor, or O2asis to 
return the device.

Problem

Possible Cause

Action to Take

The temperature 
of the air coming 
out of the can-
nula feels colder 
or warmer than 
usual

You may be using 
the device in a 
colder or warmer 
environment 
than usual, or the 
device could be 
improperly placed 
on or near a heat-
er or cooler.

Change the temperature setting on the device 
to be lower or higher than usual based on your 
surrounding environment. Also make sure that 
the device is securely positioned away from any 
sources of cold, heat or humidity. If the device 
still does not provide the proper air tempera-
ture, please contact the care provider, licensed 
distributor, or O2asis to return the device.

You are not able 
to adjust the 
heat level of the 
humidifier

There is a mal-
function with the 
control buttons, 
or the device has 
no power.

Unplug the device from the wall outlet and then 
plug back in. If that doesn’t solve the problem, 
check the connection of the power supply bar-
rel plug into the device. Make sure it is securely 
connected. If you are still not able to adjust the 
settings on the device, please contact the care 
provider, licensed distributor, or O2asis to return 
the device.

PP-19-1 REV. 15

Содержание O2asis

Страница 1: ...Personal Humidifier USER GUIDE PP 19 1 REV 15...

Страница 2: ...und to be defective within the stated warranty period Should the O2asis base unit or the replacement water well fail prematurely the replacement base unit is warranted for the remainder of the origina...

Страница 3: ...djusting the Heat Level 13 Pediatric Patients 13 Use Without Electrical Power 14 Replacing the Water Well Consumable Cartridge 15 Checking for an empty Water Well 17 Cleaning and Disinfection 18 Regul...

Страница 4: ...the power cord in a way that will avoid anyone tripping or anything otherwise interfering with it Do not position the device in a way that would make it difficult to access the power supply connection...

Страница 5: ...er Do not attempt to use with or connect to other medical equipment This device is not intended for connection to a medical pipeline This device is not intended for use with patients whose upper airwa...

Страница 6: ...ryloni trile which is known in the state of California to cause cancer For more information go to www p65warnings ca gov INCLUDED WITH THE SYSTEM The O2asis Humidifier comes with the following parts N...

Страница 7: ...wer Cord Connector The connector that accepts the male connector at the end of the power cord 6 Heated Line The heated tube attached to the Water Well cartridge that heats the oxygen to prevent water...

Страница 8: ...ns Heated Line Power Levels Water Heater Temperature Levels Water Heater power turned off when Heated Line is not detected LED Indicator Lights 3 LED Indicator Lights low medium high User Interfaces T...

Страница 9: ...e For additional reservoir durations please refer to the Performance section of this manual Note The device will detect when the Water Well is empty and will shut down The three LED lights will flash...

Страница 10: ...o prevent over filling the pump should not be used while the Water Well is disconnected from the Base Unit Confirm that the electrical port is dry and reattach the device to an electrical outlet If th...

Страница 11: ...ment Note Make sure that the Water Well consumable cartridge is securely latched to the base unit before plugging in the power cord Note The O2asis Humidifier device will continue to humidify the airf...

Страница 12: ...steps Warning Use only the cannula included with the device Use of other cannulas can degrade the performance of the device 1 Remove the cannula from its plastic bag 2 Connect the tube end to the conn...

Страница 13: ...ed oxygen flow as shown in the table below 2 Above 2lpm of oxygen flow begin your oxygen therapy on the first level of heat When the device receives power the left most LED light will illuminate signi...

Страница 14: ...in Adjusting the Heat Level sec tion of this document Condensation Regularly check for visual condensation in the cannula If excessive amounts of condensation are present reduce the heat setting to t...

Страница 15: ...age device To replace the Water Well consumable cartridge complete the following steps 1 Unplug the power adapter cord from the latch side of the device 2 Unplug the oxygen source line from the oxygen...

Страница 16: ...ging 8 Insert the non port side of the cartridge into the button side of the base and push in until the cartridge sits flush with the base 9 Swing the base latch up and snap it onto the cartridge to l...

Страница 17: ...d can become contaminated with bodily fluids If visible contamination is present or performance degradation is suspected dispose of and replace the Water Well consumable cartridge Warning Do not touch...

Страница 18: ...oap Wipe all exterior surfaces If visible soil is present use a wetted soft nylon bris tle brush to remove visible soil paying close attention to seams and crevices in the device Wipe again with dish...

Страница 19: ...ganic solvents or abrasive cleaners These types of cleaners may create toxic gasses when heated and may damage the device Warning Water fill pump is intended for single patient use only The Water Fill...

Страница 20: ...hysical Dimensions 187 x 102 x 30mm 7 4 x 4 0 x 1 2 in Not including 1118mm 44 in long heated line Weight 0 7 lbs 65lbs 295 g empty 10 95 lbs 430 g full Distilled water reservoir approx capacity 135 m...

Страница 21: ...ion This device is an US FDA cleared Class II medical device This device is considered an IEC Class II continuous operation ordinary equipment device The heated line with cannula is considered an IEC...

Страница 22: ...w Rate lpm liters per minute Leakages ml milliliters Oxygen Supply Flow Rate Range 0 5 lpm to 6 lpm Maximum Oxygen supply flow rate 6 lpm ESTIMATED HUMIDIFICATION OUTPUT Dry Oxygen Flow Rate Heat Sett...

Страница 23: ...Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the O2asis Humidifier Syst...

Страница 24: ...domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Immunity Electromagnetic Immunity The O2asis Humidifier is intended for use in...

Страница 25: ...io bands 10 V m The O2asis Humidifier is suitable for the electromagnetic environment of typical homes commercial or hospital settings DISPOSAL Base Unit This device should be disposed of separately f...

Страница 26: ...snug If the device still does not provide the proper air flow please contact the care provider licensed distributor or O2asis to return the device Problem Possible Cause Action to Take The temperatur...

Страница 27: ...ng comfort anymore please purchase a new Water Well to restore full comfort Problem Possible Cause Action to Take The device does not seem to be operating correctly You accidentally spilled water onto...

Страница 28: ...1001 New Hampshire Ave Lakewood NJ 08701 USA Toll Free 1 800 337 3762 www o2asishumidifier com...

Отзывы: