Caution: 

Hot surface. The inside of the incubation chambers can be heated to 70°C depending on the temperature

selection. Check the temperature from the display before touching the inside of the incubation chambers. Be aware

that the plate holder and the plate itself have the same temperature as the chamber they are located in.

Caution: 

Hot liquids. The temperature of the liquids inside the instrument can reach 70°C. Mishandling the plate can

cause the liquid in it to spill or splash. The spilled liquid may cause scalds or burns if the liquid comes into contact

with skin.

Chemical and biological hazards

Warning: 

Biohazard warning. The plates may contain biohazardous or chemically contaminated materials.

Warning: 

The plates are not fastened to the plate holder. Use caution when moving plates in their plate holders,

especially if they contain biohazardous material.

Caution: 

Liquid can vaporize when heated. You are responsible for checking with the liquid supplier that the

potential fumes from the vaporized liquid are not hazardous.

When operating the Trinest instrument, treat all samples and waste as potentially infectious and always wear

appropriate protective equipment, consisting of a laboratory coat, protective gloves and safety glasses.
Do not touch the labware with unprotected hands to avoid direct contact with potentially hazardous substances and to

prevent any cross contamination of the hardware.
If the Trinest instrument becomes contaminated with any sample material; it must be disinfected and cleaned as soon

as possible. See 

Routine maintenance.

Risks associated with operation and maintenance

Caution: 

Do not use plates with a lid, frame or collar around the top. Using these types of plates may interfere with

the plate detection mechanism. Three SBS standard size microplates can be loaded into one unit.

Warning: 

The plates are not fastened to the plate holder. Use caution when moving plates in their plate holders,

especially if they contain biohazardous material.

Warning: 

If the plate holder mechanism is worn out, it may spill during shaking. Spillage may result in cross-

contamination of the samples or in the contamination of the instrument and the plate holders. Do not touch the

labware with unprotected hands and avoid direct contact with potentially hazardous substances,

Warning: 

Make sure that the incubation temperature is correct when loading the plates and during the incubation.

The instrument does not give an error if the actual incubation temperature deviates from the target incubation

temperature.

Warning: 

If the instrument chambers are already heated and you insert a plate without starting the incubation cycle,

the plate incubates for as long as the temperature remains high enough. In this case, the instrument timer does not

take the incubation time into account.

Warning: 

The plate incubates for as long as the temperature remains high enough unless it is removed from the

instrument.

Caution: 

Keep your hands, head, and other body parts away from the moving parts when the instrument is in

operation.

Warning: 

Disconnect the power supply before performing any maintenance on the instrument. Switch the

instrument off from the power switch and disconnect the power supply cable from the mains.

Caution: 

Allow the incubation chambers to cool down before cleaning them.

Warning: 

Do not clean the instrument with chlorine bleach, detergents based on chlorine scouring powder,

ammonium, steel wool or agents with metallic constituents. This can cause damage to the surface of the equipment.

Notice regarding serious incidents

For a patient/user/third party in the European Union and in countries with an identical regulatory regime (IVDR; EU

2017/746/EU); if, during the use of this device or as a result of its use, a serious incident has occurred, please report

TriNEST Microplate incubator and shaker

Safety information

5

Содержание TriNEST

Страница 1: ...1296 0050 INSTRUMENT MANUAL TriNEST Microplate incubator and shaker...

Страница 2: ......

Страница 3: ...7 Unpacking 7 Environmental requirements 7 Installation 7 Functional description 9 Physical description 9 Control panel 10 Display 11 Controls 12 Operating instructions 14 Program selection 14 Paramet...

Страница 4: ...ation conditions 21 Storage conditions 21 Safety standards 21 Optional features and spare parts 21 Error handling 24 General 24 Error 1 Tmp sens n 24 Error 2 Heater n 24 Error 3 Motor belt 24 Error 4...

Страница 5: ...of microplates in accordance with the specific requirements of the different assays The instrument is semi automated and intended to be used by trained laboratory personnel The function specific diso...

Страница 6: ...se label Caution consult instruction for use instrument Biological risks Caution hot surface Consult instructions for use In vitro diagnostic medical device Catalogue number Serial number This way up...

Страница 7: ...chment part Connection Plate 2300073600 2 pcs Cleaning tool TriNEST Instrument Manual and CD including translation of the Instrument Manual Quality Control Certificate The package contains all materia...

Страница 8: ...to keep in mind when using the Trinest Microplate incubator and shaker instrument General safety Warning This equipment must be installed and used in accordance with the manufacturer s recommendations...

Страница 9: ...t Warning The plates are not fastened to the plate holder Use caution when moving plates in their plate holders especially if they contain biohazardous material Warning If the plate holder mechanism i...

Страница 10: ...ity The contact information for the manufacturer of this device to report a serious incident is Wallac Oy Mustionkatu 6 FI 20750 Turku Finland http www perkinelmer com Phone 358 2 2678 111 TriNEST Mic...

Страница 11: ...must be strong and have a flat and non slip surface Leave sufficient clearance 10 cm around the instrument for adequate air circulation Ambient temperature range during operation 15 35 C Relative humi...

Страница 12: ...re 3 Two TriNEST instruments stacked on top of each other back view Make sure the power switch is set to 0 before connecting to the mains TriNEST Microplate incubator and shaker Installation and stora...

Страница 13: ...ions and features on page 19 can be loaded at a time All three incubation chambers are controlled by the same program they cannot be programmed independently However you can stack up to three instrume...

Страница 14: ...unprotected hands and avoid direct contact with potentially hazardous substances Caution Do not use plates with a lid frame or collar around the top Using these types of plates may interfere with the...

Страница 15: ...temperature is above the target and is cooling down towards it Incubation time in hours and minutes max 48 hours Shaking speed 400 to 1200 rpm in steps of 100 or OFF no shaking Interval shake time in...

Страница 16: ...displays are rounded to the nearest half a degree Controls The controls are as follows Left and right arrow buttons used in parameter setting to step backwards or forwards through the parameters Sele...

Страница 17: ...ss it again to continue It will be disabled when incubation is completed until all plates have been unloaded Plate loaded indicators on the left of the display there are three green LEDs A LED will be...

Страница 18: ...editing press the Right or Left arrow to accept the selection and move the cursor to the next or previous parameter Program name turn the selection wheel either clockwise or counter clockwise until t...

Страница 19: ...to pre heat the incubation chambers 3 From the display check that the target incubation temperature is correct and has been reached Load plates in holders into the instrument The Start Pause button wi...

Страница 20: ...any plate just loaded including one that has been temporarily removed and then reloaded the incubation timer will reset and start from the beginning For a plate that has not been removed the timer wil...

Страница 21: ...following of the temperature during the incubation cycle Ambient this is for service purposes The temperature in the electronics compartment will be shown at the bottom of the operational display OFF...

Страница 22: ...isinfection procedure should be followed Caution Allow the incubation chambers to cool down before cleaning them Switch off the power and unplug the instrument Use disposable gloves Soak up the spill...

Страница 23: ...p Using these types of plates may interfere with the plate detection mechanism Three SBS standard size microplates can be loaded into one unit Temperature The incubation temperature can be set from 20...

Страница 24: ...ute Parameter sets Up to 20 parameters sets programs can be stored in the instrument Measurements The measurements of the TriNEST Microplate incubator and shaker instrument are as follows Physical dim...

Страница 25: ...101 IEC 61010 2 010 CAN CSA C22 2 61010 1 UL 61010 1 IEC 61326 1 IEC 61326 2 6 Optional features and spare parts The following optional features can be ordered separately from PerkinElmer Cooling blo...

Страница 26: ...e in the instrument This mechanism is subjected to wear and tear and a worn out mechanism can result in spillage especially during shaking operations In order to avoid spillage replace the plate holde...

Страница 27: ...TriNEST Microplate incubator and shaker Specifications and features 23...

Страница 28: ...he instrument and in case the instrument displays one of the below error messages reboot the instrument and if the problem persists contact authorized service person Error 1 Tmp sens n Temperature sen...

Страница 29: ...g and recovery programs vary by regulatory authority at your location Contact your local responsible body e g your laboratory manager or authorized representative for information regarding applicable...

Страница 30: ...is instrument is free from defects in materials and workmanship During the Instrument Warranty Period Wallac will provide parts and labor to repair defects in materials and workmanship Note The warran...

Страница 31: ...ion 7 Instrument description 9 Instrument operation 15 Intended purpose 1 M Maintenance 18 Measurements 20 O Operating temperature 20 Optional features 21 Optional parts 21 P Package contents 3 Pollut...

Страница 32: ...28...

Страница 33: ......

Страница 34: ...1296 9160 06 April 27 2022 Manufacturer Wallac Oy Mustionkatu 6 FI 20750 Turku Finland Tel 358 2 2678 111 Fax 358 2 2678 357 www perkinelmer com...

Отзывы: