Part number for ordering: 1296-2090
•
Plate holder
: the TriNEST incubator-shaker unit has three plate holders: one for each of the three chambers.
The plate holder not only serves as a drawer to insert the plate into the instrument, it also has a spillage
prevention mechanism whereby it raises the plate against the chamber's ceiling where the plate remains while in
the instrument. This mechanism is subjected to wear and tear and a worn-out mechanism can result in spillage,
especially during shaking operations.
In order to avoid spillage, replace the plate holders as soon as you start noticing wear in the mechanism.
Part number for ordering (spare part): 2300026300
•
Power supply
: The TriNEST incubator-shaker power supply unit to replace the PSU delivered with the instrument in
case it is broken or lost.
Part number for ordering (spare part): 1296-2040