Safety information
This section contains the safety information for the TriNEST Microplate incubator and shaker instrument.
Conventions
The following conventions are used in all system documentation:
Warning:
A warning indicates a hazardous situation or operation that, if not avoided, could result in serious personal
injury or death. Follow all stated precautions.
Caution:
A caution indicates an operation that could cause damage to the instrument or loss of data if correct
procedures are not followed.
Note:
A note emphasizes significant information in a procedure or description and alerts users to key points of interest
not related to personal injury.
In cases where a specific hazard or warning has been identified, such as an electrical or burn hazard, the related
symbol will replace the standard exclamation mark in warnings.
Warnings and precautions
This section contains the safety warnings and precautions you need to keep in mind when using the Trinest Microplate
incubator and shaker instrument.
General safety
Warning:
This equipment must be installed and used in accordance with the manufacturer's recommendations.
Installation and service must be performed by personnel properly trained and authorized by PerkinElmer.
Caution:
Misuse of electrical equipment can cause electrical injury or other hazards.
Caution:
Do not position the instrument in a way that makes it difficult to disconnect the power source. The operator
should always have unhindered access to the power switch. Both the power switch and the DC power connector are
located on the instrument's back panel.
Figure 1: Power switch and DC power inlet
Warning:
Only use the power supply delivered with the instrument.
Warning:
Maximum three stacked instruments. The attachment parts (Connection Plate 2300073600) supplied with
the instrument are required when stacking instruments on top of each other. Instrument stacking installation is only
to be performed by personnel properly trained and authorized by PerkinElmer.
TriNEST Microplate incubator and shaker