50 . STA 8000 Installation and Hardware Guide
Once Pyris Software has been configured for the STA 8000, the STA 8000 Application can be
started from Pyris Manager.
Starting the Pyris Manager
1.
Select
Pyris Manager
from the Pyris Software group in Programs on the Start menu.
The Pyris Manager is displayed on the screen.
2.
Select
STA 8000
from the Pyris Manager.
The STA 8000 control panel is displayed.
3.
Heat the sensor to approximately 150 °C to burn off any excess moisture.
Enter 150 in the box beneath the Go To Temp button
on the control panel
and click the button.
4.
After 10–15 minutes, cool the analyzer to ambient temperature.
The configuration for the STA 8000 is complete. The analyzer is ready to be calibrated.
Содержание STA 8000
Страница 1: ...STA 8000 Installation and Hardware Guide THERMAL ANALYSIS ...
Страница 4: ...Cleaning the Furnace and Sensor 61 Cleaning the Cover 61 STA 8000 Part Numbers 62 ...
Страница 5: ......
Страница 6: ......
Страница 7: ...Introduction ...
Страница 9: ...Safety and Regulatory Information ...
Страница 19: ...Safety and Regulatory Information 19 ...
Страница 20: ......
Страница 21: ...Prepare the Laboratory ...
Страница 24: ...24 STA 8000 Installation and Hardware Guide ...
Страница 25: ...Install Multiple Analyzers ...
Страница 28: ...28 STA 8000 Installation and Hardware Guide ...
Страница 29: ...Install an STA 8000 ...
Страница 54: ...54 STA 8000 Installation and Hardware Guide ...
Страница 55: ...STA 8000 Hardware ...
Страница 64: ...64 STA 8000 Installation and Hardware Guide ...