Install an STA 8000 . 43
5.
Position the sensor just above the hole in the bottom of the furnace.
The three electrical contacts must point towards the front of the analyzer.
6.
Carefully lower the sensor into the hole, keeping it in an upright position.
Do not rub the electrical contacts against the furnace insulation.
7.
When the sensor encounters resistance, release the rod.
8.
Press the top of the rod GENTLY with the tweezers and push the sensor downward a
further 2–3 mm until it does not go any further.
As the sensor makes contact with the balance mechanism, a beep will sound.
CAUTION
ATTENTION
Do not push down on any part of the sensor except the top of the rod.
Ne poussez pas sur n'importe quelle partie du capteur, sauf le haut de la
tige.
To remove the sensor:
Grip the sensor at the top of the rod with tweezers, and pull it straight upwards and out
of the furnace.
Two short beeps will sound.
NOTE: Always place the sensor in its box if it is not going to be put back in the instrument
immediately. Replace the shipping/alignment tool if the analyzer is to be moved.
Содержание STA 8000
Страница 1: ...STA 8000 Installation and Hardware Guide THERMAL ANALYSIS ...
Страница 4: ...Cleaning the Furnace and Sensor 61 Cleaning the Cover 61 STA 8000 Part Numbers 62 ...
Страница 5: ......
Страница 6: ......
Страница 7: ...Introduction ...
Страница 9: ...Safety and Regulatory Information ...
Страница 19: ...Safety and Regulatory Information 19 ...
Страница 20: ......
Страница 21: ...Prepare the Laboratory ...
Страница 24: ...24 STA 8000 Installation and Hardware Guide ...
Страница 25: ...Install Multiple Analyzers ...
Страница 28: ...28 STA 8000 Installation and Hardware Guide ...
Страница 29: ...Install an STA 8000 ...
Страница 54: ...54 STA 8000 Installation and Hardware Guide ...
Страница 55: ...STA 8000 Hardware ...
Страница 64: ...64 STA 8000 Installation and Hardware Guide ...