Fiber Optic Probe .
81
Removing the FOP
If it is necessary to remove the FOP, for example, for replacement or transportation, please
follow this procedure carefully:
1.
Ensure the attenuator cap is attached to the FOP.
2.
Switch off the instrument and remove the main power cable and, if appropriate, the
stage controller power cable.
3.
Carefully move the instrument to permit easier access to the rear panel.
4.
Remove the plug in the socket marked LED FIBER #1.
5.
Carefully remove the retained ends of the FOP by unscrewing and pulling the cables
away from the couplers perpendicularly.
Be extremely careful when removing the FOP as you can easily damage the instrument.
6.
Replace the plastic caps on the ends of the FOP that have been removed from the rear
panel.
7.
Fasten the retained screw-in caps onto the couplers on the rear panel.
8.
Replace the fiber port cover assembly if required (see
Attaching the Fiber Port Cover
Assembly
on page 49).
9.
Reconnect the power cables.
Содержание RamanStation 400
Страница 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 RamanStation 400 Series Getting Started Guide...
Страница 11: ...Safety Practices...
Страница 28: ...28 RamanStation 400 Series Getting Started Guide...
Страница 29: ...Raman History Theory and Instrumentation...
Страница 37: ...Preparing Your Facility...
Страница 42: ...42 RamanStation 400 Series Getting Started Guide...
Страница 43: ...Installing Your RamanStation 400...
Страница 71: ...Introduction to the RamanStation 400...
Страница 76: ...76 RamanStation 400 Series Getting Started Guide...
Страница 77: ...Fiber Optic Probe...
Страница 84: ...84 RamanStation 400 Series Getting Started Guide...
Страница 85: ...Using the RamanStation 400...
Страница 106: ...106 RamanStation 400 Series Getting Started Guide...
Страница 107: ...Maintenance and Troubleshooting...
Страница 118: ...118 RamanStation 400 Series Getting Started Guide...