Installing Your RamanStation 400 .
49
Attaching the Fiber Port Cover Assembly
NOTE: This section is only applicable if you have a RamanStation 400F.
The RamanStation 400F is classified as a Class 3B laser product according to the label
attached to the instrument. However if required, the RamanStation 400F can be classified as
a Class 1 laser product provided the following conditions are met:
•
No fiber optic probe is connected to the instrument;
•
The fiber optic probe cover assembly (L1320263) is attached to the instrument and
completely encloses the EXC and COL couplers on the rear panel.
A Class 1 laser product is safe under all reasonably foreseeable conditions of use and may be
used in an open laboratory environment.
To attach the fiber port cover to the rear panel, fasten the supplied screw through the cover
into the hole near the COL coupler as shown in Figure 17.
Figure 17 Attaching the fiber port cover
Screw
Содержание RamanStation 400
Страница 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 RamanStation 400 Series Getting Started Guide...
Страница 11: ...Safety Practices...
Страница 28: ...28 RamanStation 400 Series Getting Started Guide...
Страница 29: ...Raman History Theory and Instrumentation...
Страница 37: ...Preparing Your Facility...
Страница 42: ...42 RamanStation 400 Series Getting Started Guide...
Страница 43: ...Installing Your RamanStation 400...
Страница 71: ...Introduction to the RamanStation 400...
Страница 76: ...76 RamanStation 400 Series Getting Started Guide...
Страница 77: ...Fiber Optic Probe...
Страница 84: ...84 RamanStation 400 Series Getting Started Guide...
Страница 85: ...Using the RamanStation 400...
Страница 106: ...106 RamanStation 400 Series Getting Started Guide...
Страница 107: ...Maintenance and Troubleshooting...
Страница 118: ...118 RamanStation 400 Series Getting Started Guide...