Using the RamanMicro 200 Series . 81
In the unlikely event that the wavelength calibration verification should fail, follow Figure 41.
If the wavelength calibration verification fails for the microscope, this does not necessarily
mean that the performance of any other accessories attached to your instrument is also
affected.
SOP-11: Wavelength Calibration
Wavelength calibration should only be performed if the instrument has failed a wavelength
calibration verification.
1.
Ensure that the RamanMicro 200 is switched on.
If not, follow SOP-1.
2.
Attach the external neon calibration accessory (L1320220) to the rear panel of the
spectrometer module.
Push the large connector into the socket labeled CALIBRATION or CAL ACCY and
carefully attach the small cable to the COL coupler below the LED FIBER 1 connector.
Figure 42 Attaching the calibration accessory
NOTE: As wavelength calibration is not required on a regular basis, the external neon
calibration accessory is not included in the basic RamanMicro 200 package. The
external neon calibration accessory is included in the RamanMicro Calibration Kit
(L1320266), which can be purchased separately.
3.
Rotate the objective nosepiece until mid-way between two objectives.
4.
Cover the microscope with a dark cloth or, if fitted, close the light-proof enclosure doors
and ensure that caps are placed on any external fiber optic probes or accessories, to
minimize stray light from interfering with the calibration.
5.
Select the Dark Field (DF) switch position on the microscope and switch off the
illumination.
6.
Select
Raman Instrument
from the Setup menu.
The Setup Instrument tabs are displayed.
7.
Select the Setup Instrument Advanced tab and click
Toolbox
.
The Adjustments Toolbox dialog is displayed.
8.
Select
.
The Wavelength Calibration dialog is displayed.
Содержание RamanMicro 200 Series
Страница 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 RamanMicro 200 Series Getting Started Guide...
Страница 11: ...Safety Practices...
Страница 27: ...Raman History Theory and Instrumentation...
Страница 33: ...Preparing Your Facility...
Страница 37: ...Installing Your RamanMicro 200 Series...
Страница 57: ...Using the RamanMicro 200 Series...
Страница 89: ...Maintenance and Troubleshooting...
Страница 98: ...98 RamanMicro 200 Series Getting Started Guide...