34 .
RamanMicro 200 Series Getting Started Guide
Specifications
Spectrometer
Microscope
Stage Controller
(for motorized XYZ
stage only)
Weight
20 kg
20 kg
2.8 kg
Height
161 mm
520 mm
90 mm
Width
530 mm
370 mm
220 mm
Depth
600 mm
430 mm
184 mm
Power Supply
100–230 V ±10%,
50–60 Hz
100–120 V/220–240 V
50–60 Hz
100–240 V,
50–60 Hz
Primary Fuse
–
See Olympus manual
1.6 A T (time-lag)
Laser Class
Class 3B
–
Operational
Temperature
15 °C – 30 °C
Recommended
Temperature
20 °C – 25 °C
If the electricity supply does not conform to these specifications, please consult PerkinElmer
prior to installation. If the supply of power is erratic please use an uninterruptible power
supply (UPS). Incorrect shutdown, power fluctuations or brown-outs may damage the
spectrometer.