RamanMicro 200 Series Getting Started Guide
Switching on the Instrument
NOTE: Do not connect the USB from the alignment camera to the PC until after the software
Installing the Instrument in Spectrum Software
on page 49.
1.
Insert the key into the laser power
keyswitch on the front panel of the spectrometer
module and ensure that it is turned to the ARM position.
The RamanMicro 200 is supplied with two keys. One should be placed in the instrument
and the other should be stored in a safe place. Replacement keys (L1323503) can be
obtained from PerkinElmer. A laser safety officer or lab manager can control use of the
instrument with these keys.
2.
Connect the instrument power supply unit (L1321802) to the RamanMicro 200
spectrometer
Main Power
socket and then switch on.
Do not connect to the Aux Power connector.
3.
Connect the power cable from the microscope to the mains supply (Figure 17), and then
switch on the microscope.
4.
If you have a motorized stage, switch on the stage controller box (Figure 15).
Содержание RamanMicro 200 Series
Страница 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 RamanMicro 200 Series Getting Started Guide...
Страница 11: ...Safety Practices...
Страница 27: ...Raman History Theory and Instrumentation...
Страница 33: ...Preparing Your Facility...
Страница 37: ...Installing Your RamanMicro 200 Series...
Страница 57: ...Using the RamanMicro 200 Series...
Страница 89: ...Maintenance and Troubleshooting...
Страница 98: ...98 RamanMicro 200 Series Getting Started Guide...