This document describes the dimensions, installation and IT requirements of the
Opera Phenix Plus
®
Plus High Content Screening System.
Order Review
Please review this order. Record any discrepancies between the PerkinElmer order
and your purchase order, along with any agreements or commitments made by your
PerkinElmer sales representative that are NOT listed on the order. Let your sales
representative know about these discrepancies and/or commitments with your Site
Readiness confirmation.
Notice
The last page of this document needs to be filled and sent back to
your PerkinElmer Sales Representative via fax/email. The Opera
Phenix Plus will not be shipped until confirmation fax/email has been
received by PerkinElmer.
The Way of Your Opera Phenix Plus
1. Manufacturing of main unit: To achieve maximum robustness, the basic unit of
your Opera Phenix Plus is manufactured within a small series.
2. You receive the Site Readiness document.
3. Customization: Your Opera Phenix Plus is equipped with the features you
ordered.
4. Internal Acceptance Test (IAT)
To assure that all components of your Opera Phenix Plus work within the
specifications, the performance of the whole system is checked thoroughly.
5. Packaging: All devices and accessories are packed in specially designed
transport boxes.
6. You send the Site Readiness confirmation form to PerkinElmer.
7. Pick up: The packed Opera Phenix Plus is picked up by our co-operating shipper.
8. Delivery to you: Direct transport by truck or transport by air cargo.
9. Setting up by Service: Your system is unpacked, installed and set up by
PerkinElmer Service.
10. Site Acceptance Test (SAT)
Our Field Service engineer will test the instrument in order to ensure that its
performance meets manufacturing specifications.
Note: The SAT procedure is not intended as a substitute for an IQ/OQ
qualification process.We will not provide an IQ/OQ qualification document for the
Opera Phenix Plus instrument.We recommend a PerkinElmer service contract
for keeping up and confirming of proper functionality of the instrument during its
life time.
11. Familarization: After an introduction and device familarization by our Field
Service engineer you can start working with your new Opera Phenix Plus.
12. On-site user training: We offer an additional advanced user training upon request.
1 Introduction
Opera Phenix Plus – Site Readiness
7
Содержание OPERA PHENIX PLUS
Страница 1: ...OPERA PHENIX PLUS Site Readiness...
Страница 4: ......
Страница 5: ...Opera Phenix Plus Site Readiness 5...
Страница 6: ...1 Introduction 1 Introduction 6 Opera Phenix Plus Site Readiness...
Страница 8: ......
Страница 9: ...2 Requirements 2 Requirements Opera Phenix Plus Site Readiness 9...
Страница 27: ...3 IT Policy 3 IT Policy Opera Phenix Plus Site Readiness 27...
Страница 43: ...4 Confirmation Fax Email 4 Confirmation Fax Email Opera Phenix Plus Site Readiness 43...
Страница 45: ......