l
The LTSC servicing model prevents Windows
®
10 Enterprise LTSB/LTSC*
devices from receiving the usual feature updates and provides only quality
updates to ensure that device security stays up to date.
* LTSB = Long-term Servicing Branch (recently renamed to LTSC, but still
present in the name of previous releases)
l
https://docs.microsoft.com/en-us/windows/deployment/update/waas-overview
How to get security updates?
l
If the device PC is connected to the internet, you will be notified when new
security updates are ready to be downloaded. After going to Windows Update,
you can download and install the updates.
l
Updates which require a restart will not take effect until the device PC is
restarted manually. Please check regularly for corresponding notifications and
restart the PC.
l
If the update behavior is not as described above, please open the Group Policy
Editor (run "gpedit.msc"), verify the update settings and reset to the configuration
below, if necessary. Administrator rights are required for this step. The "Windows
Update" section can be found in this branch: "Computer Configuration –
Administrative Templates – Windows Components – Windows Update".
o
"Configure Automatic Updates": E option 2 selected ("2 – Notify for
download and notify for install")
o
"No auto-restart with logged on users for scheduled automatic updates
installations": Enabled
However, in the case of critical updates, Windows will ignore these settings and
still restart automatically. You cannot prevent that.
l
Check for Windows updates regularly and make sure that all available security
updates have been downloaded and installed before starting a measurement. If
you do not install pending updates for a long time, Windows will install them
automatically and the restarting of your computer may interfere with
measurements.
Can I upgrade to a new LTSC release to get new features?
No, the license is only valid exactly for Microsoft Windows
®
10 IoT Enterprise
2016/2019 LTSC (Version 1607/1809).
3.2.3
Microsoft SQL Server
l
Harmony 5.0 is using the MSSQL2012SP3. This is the only supported SQL
Server version for Harmony 5.0.
3 IT Policy
36
Opera Phenix Plus – Site Readiness
Содержание OPERA PHENIX PLUS
Страница 1: ...OPERA PHENIX PLUS Site Readiness...
Страница 4: ......
Страница 5: ...Opera Phenix Plus Site Readiness 5...
Страница 6: ...1 Introduction 1 Introduction 6 Opera Phenix Plus Site Readiness...
Страница 8: ......
Страница 9: ...2 Requirements 2 Requirements Opera Phenix Plus Site Readiness 9...
Страница 27: ...3 IT Policy 3 IT Policy Opera Phenix Plus Site Readiness 27...
Страница 43: ...4 Confirmation Fax Email 4 Confirmation Fax Email Opera Phenix Plus Site Readiness 43...