2.3
Dimensions
The Opera Phenix Plus system consists of Opera Phenix Plus device, Harmony PC
and monitor/keyboard cart. The setup and installation options are described in the
following chapters.
Notice
The Opera Phenix Plus is sensitive to vibrations. Blurred images can
be the consequence.
l
Put the Opera Phenix Plus on a very stable bench or table.
l
Do not put other devices on the same table which cause
vibrations (e.g. a centrifuge).
l
Avoid any shocks or vibrations during measurements.
Component
Size [cm]
(W×D×H)
Size [in]
Weight [kg]
Weight [lb]
Opera Phenix Plus
Plus
Instrument height
(sample lid opened)
138 × 66
85
55 × 26
34
270-300
595-662
Opera Phenix Plus
Plus with optional
pipettor
Instrument height
(housing
closed/opened)
138 × 66
95/138
55 × 26
38/55
320-350
705-772
Harmony PC
22 × 53 × 44
9 × 21 × 18
17
38
Monitor/Keyboard Cart
(keyboard tray in/out)
67 × 58/74
26 × 23/29
41
91
Opera Phenix Heavy
Duty Lab Cart
(optional)
140 × 95* × 79
56 × 38* × 31
170
375
The table will come on 4 feet, 50 cm² per single foot.
plate::handler FLEX
(optional)
214 × 95 × 181
(overall)
84 × 38 × 71
135
298
plate::handler FLEX
and incubator
(optional)
274 × 95 × 181
(overall)
108 × 38 × 71
225
496
* Rear shelf can be removed for transport, thereby depth is reduced to 80 cm (32 in).
2 Requirements
12
Opera Phenix Plus – Site Readiness
Содержание OPERA PHENIX PLUS
Страница 1: ...OPERA PHENIX PLUS Site Readiness...
Страница 4: ......
Страница 5: ...Opera Phenix Plus Site Readiness 5...
Страница 6: ...1 Introduction 1 Introduction 6 Opera Phenix Plus Site Readiness...
Страница 8: ......
Страница 9: ...2 Requirements 2 Requirements Opera Phenix Plus Site Readiness 9...
Страница 27: ...3 IT Policy 3 IT Policy Opera Phenix Plus Site Readiness 27...
Страница 43: ...4 Confirmation Fax Email 4 Confirmation Fax Email Opera Phenix Plus Site Readiness 43...
Страница 45: ......