2.8.4
Automated Incubator (plate::handler system)
In the case an optional automated cell incubator is included in the delivery of the
plate::handler system the following additional site readiness requirements need to be
considered:
Requirement
Cell Incubator Extension:
Liconic STX 44-IC
Electrical
100V / 50Hz; 115V / 60Hz or 230V / 50Hz; switchable
Power consumption
600 W
CO
2
supply
Nominal
pressure
0.5 bar
Maximum
pressure
8 bar
Maximum
flow rate
14 liter/minute (3.7 gallons/minute) (safety relevant, see note)
Tubing
Use 4x6 mm PUR, PVC, Neoprene or Nylon tube for the low
pressure connection. Do not use silicon or natural rubber tubing!
Size (W/D/H in mm)
485 / 535 / 1203
Weight
88 kg
Notice
It is your responsibility to ensure that the gas flow will not exceed the
limit of the maximum gas flow. In case of leakages this could be
safety relevant!
The cell incubator will be connected to the workstation table and can be fixated on the
lab floor by using two foot clamps.
2 Requirements
26
Opera Phenix Plus – Site Readiness