NOTE
: Keep the amount of liquid coolant to the level where the indicator is located between the ‘L’ (low) and ‘H’
(high) mark in the scale on the left side of the Peltier Controller. When the coolant is running short for the
operation, the LED of Low Water Level changes to red with an alarm sounding.
CAUTION
ATTENTION
The liquid coolant must be used as refrigerant. Do not use any w ater (tab
w ater, DI, etc). In case that the water is used as refrigerant, Peltier Controller’s
components might be corroded and its performance may be deteriorated.
Use any kind of liquid coolant in normal grade as is available (e.g. liquid coolant
including Distilled Water 70 wt%, Propylene Glycol 27 wt%, Additives 3 wt%).
Le liquide de refroidissement doit être utilisé comme réfrigérant. N'utilisez
jamais d'eau (eau du robinet, eau déminéralisée, etc.) L'eau risque en
effet de corroder les composants du contrôleur Peltier et de réduire les
performances de ce dernier.
Utilisez un liquide de refroidissement du commerce de qualité normale (par
exemple, un liquide de refroidissement à base d'eau distillée à 70 % M/V, de
propylène glycol à 27 % M/V et d'additifs à 3 % M/V).
4.
Connect the power cord and the communication cable of the Lambda 465.
NOTE:
Make sure that the instrument is turned off when the accessory is installed.
5.
Take apart the existing cell holder in the sample compartment.
Figure 10 Taking apart the existing cell holder