09931238D
21
Measurement
NOTE
: Start the sample measurement after warming up the system for at least 20 minutes.
This accessory can be operated in Temperature Based Kinetics in the Kinetics mode and Thermal
Denaturation in the Bio analysis mode.
Temperature Based Kinetics
1.
Launch UV Lab software. When the window below is appeared, select Experiment Type and click
OK.
2.
The following dialog box will be displayed.
3.
Setup test parameters as follows:
a.
Edit Heating (Cooling) Rate (°C/min): Set the heating (cooling) temperature rate in each
temperature range. Sample is measured whenever it reaches the set temperature by the rate
(°C/min), and the sample is also measured at the Start Temperature and the End Temperature.
Click on Insert and Delete to edit the temperature range and the rate.