09931238D
22
- Start Temperature: Enter the start temperature for the measurement.
- End Temperature: Enter the end temperature for the measurement.
- Rate (°C/min): Enter the heating (cooling) rate.
b.
Holding Time: Enter the holding time before sample measurement after it reaches the each set
temperature.
c.
Rate calculation Type: Select the rate calculation type. See V-3 Kinetics Mode of UV Lab
Software User Guide for more information.
d.
Calculation Start Temperature: Enter the start temperature for the calculation.
e.
Calculation End Temperature: Enter the end temperature for the calculation.
NOTE:
Calculation Start Temperature and Calculation End Temperature must be entered within Start
Temperature and End temperature range.
4.
Select Wavelength Range and enter the wavelength range for the measurement.
5.
After setting parameters for Experiment Setup, Baseline Correction and Temperature
Based Kinetics, select OK.
6.
Select Single Cell Peltier in the Accessory Type on the main screen of the software.
7.
Select setup button (pencil icon) in the Experiment Setup on the main screen of the
software.