Option: This function is used to adjust temperature. It is only used for the manufacturing
process, so do not modify the values.
9.
After completing the parameter setup, select OK. Then the LED for On-LINE is turned on and it
will start heating up or cooling down to the set temperature in the Setup Single Cell Peltier.
10.
Insert the blank solution into the cell holder. Select Blank, and then the blank is measured when
it reaches the set Start Temperature.
NOTE:
To monitor the probe temperature, the probes should be immersed in the sample, or to use the block
temperature for monitoring, the cell lid should be closed tightly
11.
Remove the blank solution from the sample cell holder and insert the sample solution into the
sample cell holder. Select Sample and then the measurement will be started after the
temperature reaches the target temperature.