Lambda 365 Installation . 33
Environmental Requirements
This system is for indoor use only.
Requirement
Description
Indoor use
Altitude up to 2000 m (6,500 ft.)
Pollution degree 2
Dimensions
491 mm (W) × 538 mm (D) × 258 mm (H)
[19.3 in. (W) 21.1 in. (D) 10.1 in. (H)]
Weight
23.5 kg (52 lbs)
Electrical supply
AC 100-240V, 50/60Hz, 110W
Operating Temperature
15 - 35
℃
(59 - 95°F)
Operating humidity
Less than 80% relative, non-condensing
Mains Supply Voltage
Functions
± 10%
Computer Requirements
Requirement
Description
Hardware
Processor : Intel® Core 1.5 GHz or faster
RAM : At least 1GB
Hard disk : 50GB with 1GB free
Input devices : Mouse and keyboard
Monitor : 1024x768 (minimum)
Media : CD ROM Drive
Port : USB port for the data acquisition
Operating System
Microsoft
®
Windows 7
Output device
Microsoft
®
Windows compatible printer
Содержание Lambda 365
Страница 1: ...LAMBDA 365 Users Guide UV SPECTROMETERS ...
Страница 4: ...4 Lambda 365 Users Guide ...
Страница 5: ...Introduction 5 Introduction ...
Страница 19: ...Safety Practices 19 Safety Practices ...
Страница 30: ...30 Lambda 365 Users Guide ...
Страница 31: ...Lambda 365 Installation 31 Lambda 365 Installation ...
Страница 35: ...UV Express Software Installation 35 UV Express Software Installation ...
Страница 37: ...Interface Setup 37 3 UV Express InstallShield Wizard starts Select Next ...
Страница 41: ...Interface Setup 41 Interface Setup ...
Страница 43: ...Interface Setup 43 5 Double click USBCom CDM_20828 6 Select Extract ...
Страница 46: ...46 Lambda 365 Users Guide ...
Страница 47: ...Setting USB Serial Ports 47 Setting USB Serial Ports ...
Страница 48: ...48 Lambda 365 Users Guide Procedure 1 Select Computer Properties 2 Click on Device Manager ...
Страница 53: ...Operating Procedure 53 Operating Procedure ...
Страница 55: ...Accessories 55 Accessories ...
Страница 62: ...62 Lambda 365 Users Guide ...
Страница 63: ...Accessories 63 Lamp Replacement ...