58 . High-Performance Lambda Spectrometers Hardware Guide
To install the single-cell holder in the sample compartment:
1.
Orientate the holder so that the word LAMBDA is toward the front of the sample
compartment (see Figure 27).
2.
Lower the holder so that the two locating holes slip onto the two studs on the two
locating pins on the plinth in the sample compartment.
The cell holder fits one way round only.
Figure 27 Baseplate of the high-performance Lambda spectrometer sample
compartment
3.
Move the milled posts a little to locate the threaded holes in the baseplate and then
tighten the milled posts.
You can lead tubes and connecting cables for accessories into the sample compartment
through either the tube ports, located at the front of the sample compartment, or the
opening located at the left rear of the sample compartment, under the cover plate.
When not in use, you should always insert the caps into the tube ports and keep the
cover plate fixed in place.
The cover plate is fixed in place by a retaining screw.
Aligning the Single-Cell Holder
To carry out the coarse alignment of the single-cell holder:
1.
Open the sample compartment cover.
2.
Fill the cells with a low-absorbing solvent (deionized water or ethanol).
3.
Insert one cell into the sample-cell holder and one into the reference-cell holder.
Make certain that the cell is pushed down fully.
NOTE: The alignment procedure is for a given cell in a given holder. After alignment, the cell
should always be used in the same holder.
Cover plate
Sample holder
and plinth
Drain hole
Locating pin
Retaining screw
Tube ports
Содержание Lambda 1050
Страница 1: ...HIGH PERFORMANCE LAMBDA SPECTROMETERS Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 11: ...Safety Practices ...
Страница 25: ...System Description ...
Страница 32: ...32 High Performance Lambda Spectrometers Hardware Guide ...
Страница 33: ...Installing your Lambda Spectrometer ...
Страница 54: ...54 High Performance Lambda Spectrometers Hardware Guide ...
Страница 55: ...Using the Spectrometer ...
Страница 64: ...64 High Performance Lambda Spectrometers Hardware Guide ...
Страница 65: ...Maintenance ...
Страница 76: ...76 High Performance Lambda Spectrometers Hardware Guide ...
Страница 77: ...Appendices ...
Страница 81: ...Appendices 81 T temperature sensor 52 U unpacking 35 UV radiation safety 16 W WEEE directive 78 ...
Страница 82: ...82 High Performance Lambda Spectrometers Hardware Guide ...