46 . High-Performance Lambda Spectrometers Hardware Guide
Slits:
There are no user-defined settings in the Slits calibration.
Zero Percent T:
There are no user-defined settings in the Zero Percent T calibration.
5.
Click
Calibrate
.
The calibrations are performed in the order listed in the table. Messages below the table
show the progress of the calibration.
If you are performing the Zero Percent T calibration on a Lambda 650/750, you will be
prompted to place the magnetic beam blocks supplied over the ports in the sample
compartment wall.
When a calibration has passed, a green check mark is displayed in the first column of
the table. If the test fails, a red cross is displayed. In such an event please contact your
local PerkinElmer Service Engineer.
6.
When the calibration has finished, click
Close
to close the dialog.
The date and time of the calibration are recorded in the instrument event log.
NOTE: It is not possible to print the calibration results.
Your instrument is now ready to be used.
Содержание Lambda 1050
Страница 1: ...HIGH PERFORMANCE LAMBDA SPECTROMETERS Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 11: ...Safety Practices ...
Страница 25: ...System Description ...
Страница 32: ...32 High Performance Lambda Spectrometers Hardware Guide ...
Страница 33: ...Installing your Lambda Spectrometer ...
Страница 54: ...54 High Performance Lambda Spectrometers Hardware Guide ...
Страница 55: ...Using the Spectrometer ...
Страница 64: ...64 High Performance Lambda Spectrometers Hardware Guide ...
Страница 65: ...Maintenance ...
Страница 76: ...76 High Performance Lambda Spectrometers Hardware Guide ...
Страница 77: ...Appendices ...
Страница 81: ...Appendices 81 T temperature sensor 52 U unpacking 35 UV radiation safety 16 W WEEE directive 78 ...
Страница 82: ...82 High Performance Lambda Spectrometers Hardware Guide ...