30 . High-Performance Lambda Spectrometers Hardware Guide
Figure 9 Optical system in the Lambda 1050 spectrometer lamp compartment
Second, the automatic internal sample beam and reference beam attenuators have the
following possible values: 0%, 0.1%, 1% and 100%.
Third, the detector compartment has the Three Detector Module fitted as standard. A
photomultiplier tube (PMT) detector is used in the UV/Vis range (175–860 nm). In the NIR
range you will have either a narrowband or a wideband InGaAs detector (860–1800 nm or
860–2500 nm, respectively) and a lead sulfide (PbS) detector (860–3300 nm).
NOTE: The Three Detector Module is installed as standard in the Lambda 1050, but can also
be purchased as an accessory for the Lambda 950. For more details see the
Three
Detector Module User’s Guide (L6050010) on the High-Performance Lambda
Spectrometers User Manuals CD (L6050009).
Figure 10 Optical system in the Lambda 1050 spectrometer detector compartment
The radiation passing alternately through the sample and reference beams is reflected by
mirrors M11, M12, M13, and M11′, M12′, M13′, respectively of the optics in the detector
assembly onto the appropriate detector.
M11′
M11
M14
M15
M13′
M13
M12
M12′
PbS
PMT (below PbS)
InGaAs
DL
HL
M1
M0
M2
Содержание Lambda 1050
Страница 1: ...HIGH PERFORMANCE LAMBDA SPECTROMETERS Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 11: ...Safety Practices ...
Страница 25: ...System Description ...
Страница 32: ...32 High Performance Lambda Spectrometers Hardware Guide ...
Страница 33: ...Installing your Lambda Spectrometer ...
Страница 54: ...54 High Performance Lambda Spectrometers Hardware Guide ...
Страница 55: ...Using the Spectrometer ...
Страница 64: ...64 High Performance Lambda Spectrometers Hardware Guide ...
Страница 65: ...Maintenance ...
Страница 76: ...76 High Performance Lambda Spectrometers Hardware Guide ...
Страница 77: ...Appendices ...
Страница 81: ...Appendices 81 T temperature sensor 52 U unpacking 35 UV radiation safety 16 W WEEE directive 78 ...
Страница 82: ...82 High Performance Lambda Spectrometers Hardware Guide ...