Advanced User Settings
38
P/N CLS155330 Rev. A
FlexDrop iQ User Manual
PerkinElmer, Inc.
The following settings can be adjusted:
•
Source plate type
: Choose the source plate that will be used for
calibration.
•
Set dosing energy
: Choose the pressure to apply for calibration.
Follow the instructions and click
Check calibration
to start the measurement.
CAUTION
Use only a calibrated pipette for re-calibration of the device!
ATTENTION
Utilisez une pipette étalonnée uniquement pour le réétalonnage de
l’appareil !
The FlexDrop iQ counts the number of droplets dispensed and calculates the
deviation to the expected number of droplets. Clicking
Update calibration
moves the current FlexDrop iQ slider values by the offset of the deviation
previously calculated (see Figure 32).
Figure 32. Dispense head calibration process (step 1-4). (1) Current
slider values before re-calibration. (2) Calibration protocol. (3)
Confirmation to adjust slider values according to measurement. (4)
Automatically adjusted values after measurement