Setting Operating Parameters for Immediate Change
Using Manual Mode
3 - 7
Press the right arrow key or the
key to move to the RSP/PW field. Depending
on the detector’s configuration, either RSP or PW is displayed for this field. You use
this field to select a value for the response time or peak width at half height. To change
the value, press the corresponding softkey. For more information on which value to
select, see the “Detector Response Time” topic in Chapter 8, Principles.
Press the right arrow key or the
key to set the RSP and move to the TE1 field.
enter
enter
4
The relay event may be either O, C, or M. These letters indicate the following:
O: Open - The relay opens immediately.
C: Close - The relay closes immediately.
M: Momentary - The relay closes for about two seconds, then opens.
If the previous state of the relay is Close, this action may not cause the desired opera-
tion. Momentary closures should be preceded with an Open event.
5
Press the cursor arrow key or the
key to set TE1 and move to the TE2 field.
Select a value for TE2. The relay event may be either O, C, or M, as explained in Step
5.
enter
6
Содержание 200 UV/VIS Series
Страница 1: ...Series 200 UV VIS Detector User s Manual...
Страница 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Страница 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Страница 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Страница 63: ...Making Input Output Connections Installation 2 13 Figure 2 5 Connecting a Wire to the Modular Connector...
Страница 156: ...Maintenance Troubleshooting Methods 7 26...
Страница 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Страница 176: ...Index In 6...