Instrument Specifications
Getting Started with the Detector
1 - 21
Control outputs
4 SPDT relays, 1 A @ 48 V (max); Relay TE 2 is used to control
an external recorder for scanning.
RUN OUT: momentary (approximately 2 seconds) contact clo-
sure the instant a method starts. Used to “daisy chain” modules
together.
RDY: signals that the unit is ready to run.
TE1 and TE2: programmable outputs.
Control inputs
EXT RUN: same as pressing the run front panel key.
EXT STOP: same as pressing the reset front panel key.
AZ: same as pressing the AZ front panel key.
External Dimensions
340 mm wide, 170 mm high, 495 mm deep. (13.4” W x 6.7” H x
19.5” D, Volume = 1750 c.i.)
Weight
25 lbs. 11.4 Kg.
Power Requirements
100/120/220/230/240 VAC, 50/60 Hz, 130 VA max.
Optional Configurations
Microbore and preparative flowcells, Tungsten Lamp
Specification
Description
Содержание 200 UV/VIS Series
Страница 1: ...Series 200 UV VIS Detector User s Manual...
Страница 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Страница 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Страница 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Страница 63: ...Making Input Output Connections Installation 2 13 Figure 2 5 Connecting a Wire to the Modular Connector...
Страница 156: ...Maintenance Troubleshooting Methods 7 26...
Страница 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Страница 176: ...Index In 6...