background image

•  COMMAND K 

Return 4 bytes. For example, when sends command "K" to meter, it will 

return "3","0","1", ASCII(13) .

•  COMMAND D

Return data of main window.

Range: T1, T2, T1-T2 (7bytes), the unused characters is left as space 

ASC(13). 

4. OPERATION INSTRUCTIONS

P.10

RS232 command

K(ASC 4BH)
D(ASC 44H)

B(ASC 42H)

S(ASH 53H)
H(ASC 48H)
T(ASC 54H)
M(ASC 4DH)
N(ASC 4EH)

R(ASC 52H)
C(ASC 43H)
A(ASC 41H)

Function

Ask for model No.
Ask for main display 

Range, Data, Unit
Ask for secondary 

display Range, Data, 

Unit
Ask Status
Hold button
TIMER button
AVG/MAX/MIN button
Exit AVG/MAX/MIN 

mode
REL button
C/F button
Inquire all encoded 

data

Remarks

Send 4 bytes
Send 22 bytes

Send 22 bytes

Send 13 bytes

Send encoded 8 

byte

Содержание TC0301

Страница 1: ...Datalogger Thermometer TC0301 Instruction Manual...

Страница 2: ...1 Power Up 4 2 Connecting the Thermocouples 4 3 Selecting the Temperature Scale 4 4 Data Hold Operation 4 5 T1 T2 T1 T2 Display Control 4 6 Relative Operation 4 7 MAX MIN AVG Operation 4 8 Auto Power...

Страница 3: ...ype thermocouples NUMERICAL DISPLAY 4 digital liquid crystal display MEASUREMENT RANGE 200 C 1370 C 328 F 2498 F RESOLUTION 200 C 200 C 0 1 C 800 C 1370 C 1 C 200 F 200 F 0 1 F else 1 F MAXIMUM VOLTAG...

Страница 4: ...ement the accuracy is 0 5 T1 T2 reading 2 C or 0 5 T1 T2 reading 2 F TEMPERATURE COEFFICIENT For ambient temperatures from 0 C 18 C and 28 C 50 C for each C ambient below 18 C or above 28 C add the fo...

Страница 5: ...Carrying Case Instruction Menu OPTION Soft Ware Package Program RS232 Connection Cable AC Adapter POWER REQUIREMENT 9 Volt Battery NEDA 1604 or JIS 006P or IEC6F22 BATTERY LIFE Approx 100hrs with alka...

Страница 6: ...e display data is being hold The Maximum value is now being displayed The Minimum value is now being displayed The Average value is now being displayed The reading is now under Relative Mode The Batte...

Страница 7: ...ensor connector LCD display ON OFF button HOLD button T1 T2 T1 T2 control button Relative readout button MAX MIN Average control button C F control button Offset calibration screw 11 Digital output co...

Страница 8: ...a hold operation by pressing HOLD button again When the meter is under Data Hold operation the REL and C F button are disabled 4 5 T1 T2 T1 T2 DISPLAY CONTROL One may select T1 T2 or T1 T2 to show on...

Страница 9: ...her This means that all these data is updated in the memory and the reading is the present temperature One may press to circulate the display mode among these options When the meter is under operation...

Страница 10: ...orming offset Calibration with thermal stabled ice water through VR1 will give a very good calibration result 4 11 DIGITAL OUTPUT The Digital Output is a 9600bps N 81 serial interface The RX is a 5V n...

Страница 11: ...RS232 command K ASC 4BH D ASC 44H B ASC 42H S ASH 53H H ASC 48H T ASC 54H M ASC 4DH N ASC 4EH R ASC 52H C ASC 43H A ASC 41H Function Ask for model No Ask for main display Range Data Unit Ask for secon...

Страница 12: ...n mode HOLD MAX REL if the mode is not entered the related characters will be left as space For example when the meter is under MAX display the meter will return MAX COMMAND T Equivalent to one pushin...

Страница 13: ...C 0 F 3nd BYTE bit0 1 main window value is OL 0 not OL bit1 1 main window value is minus 0 main window value is plus bit2 1 4th byte and 5th byte represent 0 4th byte and 5th byte represent bit3 1 su...

Страница 14: ...window is T2 11 Main window is T2 sub window is T1 4th BYTE first two BCD code of main window value 5th BYTE last two BCD code of main window value 6th BYTE first two BCD code of sub window value 7th...

Страница 15: ...r products and services please send us an email cs perfectprime com For B2B or project based application please send an email sales perfectprime com FOR MORE INFORMATION ABOUT PERFECTPRIME PLEASE VISI...

Страница 16: ...HORMES LIMITED cs perfectprime com G F UNIT 3 61 GLENTHORNE ROAD LONDON W6 0LJ UNITED KINGDOM 44 203 7695377 Retailer Email Address Telephone...