For setting “Type of Device” in the “Settings” panel (Menu—“Edit”--”Settings”),
select “R-100(PTL) / R-300(PTL)” for “R-300 PowerTopoLite” and other setting should be as
follows.
Please note that these settings should be common with R-2500NS's. And if the
selection of “Type of Device” is not correct it may result in missing some data.
R-2500
→
PC(DL-01)
Bits per second:
1200 (1200-9600 for sending “POLAR DATA”)
Databits:
8
Parity:
None
Stop bits:
1
PC(DL-01)
→
R-2500
Bits per second:
1200
Databits:
8
Parity:
None
Stop bits:
1
c) Note for the Memory capacity.
Data transfer failure from DL-01 to R-2500NS may cause reduction of memory capacity.
If memory capacity becomes less, please back up required data first, and then initialize
coordinates data.
To initialize coordinates data, turn on the instrument while pressing [F2]+[F5]+[ON/OFF],
and take your finger off from [ON/OFF] again.
After you see the message “COORD. DATA INITIAL”, press [F5]. Then the message
“Please wait” is displayed. When it is completed, the panel of Electronic Vial is displayed.