118
Select the file you desire to read, then press [ENT]
In this case, new job file is created in the internal
memory of the instrument and the transferred data is
recorded in it.
NOTE:
File name is limited to a maximum of 12 characters. If the file name is more than
13 characters, it is
not put in the list of the file that can be read.
NOTE:
Error Message
• Duplication of the file name.
If the same file name exists in the instrument, an error message [The file name isn’t correct]
is displayed. Change the file name and try again.
• Incorrect format.
If the format is incorrect, an error message [Format is not correct] is displayed.
Check the format of the coordinate data and try again to read the text file.
(For more details on “Format”, refer to “13. INPUT/OUTPUT”.)
• Maximum number of job files stored in the internal memory is exceeded.
The maximum number of job files stored in the file is 50. If you try to read a new file while
50 job files already exist, an error message [limited number of files ] is displayed.
Delete unnecessary job files by pressing [4. DELETE] in [FILE MANAGEMENT].
(For more details, refer to 3.4 Deletion of a Job name.)
• The remaining capacity of the internal memory becomes less than 1 MB.
You cannot create a file when the remaining capacity is not enough and an error
message [Space capacity is short] is displayed. Make sure that the remaining capacity is
more than 1MB, then try again to read the text file.
NOTE:
Warrning
• The number of data exceeds 3,000 points.
The maximum number of points stored in the internal memory per job is 3,000 points. If it
reaches 3,000 points during reading the text file, an error message ["3000 point over!"] is
displayed. In this case, up to the 3,000
th
point will be read, but after that no more.