4. User
Maintenance/Service
4.1 Fuse
Replacement
If the fuse is suspected of being defective, it should be inspected and, if necessary, replaced. To inspect or replace
the fuse, please perform the following steps:
(1)
Disconnect the AC line cord from the unit to reduce electrical shock hazard.
(2)
Remove the fuse by unscrewing the cap from the body of the fuseholder. Test the fuse for electrical
continuity with any ohmmeter.
(3)
If the fuse is found to be open, replace with a replacement fuse as specified in the following table:
AC
Line
Voltage Fuse
Rating
110/120
V
AC
1
A/250
V
220/240
V
AC
0,5
A/250
V
(4)
Replace the fuse in the fuseholder cap and re-install.
(5)
Reconnect the AC power cord.
Note: Use of any fuse other than the one specified may cause damage to the unit, pose a severe fire
hazard, and will void the warranty.
4.2
In case of difficulties
This function generator has been designed to be accurate, reliable, and easy-to-use. However, it is possible that
you may experience difficulties during operation. If there appears to be any kind of problem during use of the unit,
please perform the following steps to help determine the cause:
(1)
Re-read the operating instructions. It is very easy to inadvertently make mistakes in operating procedure.
(2)
Remove and test the fuse. The function generator will not function with an open fuse.
If the preceding two steps fail to resolve the problem, please refer to the "Obtaining Service" section.
Note: Attempted Repair, modifications, or tampering by unauthorized personnel will void the warranty.
5. Specifications
Waveforms
Sine, Square, Triangle, Ramp, TTL Pulse, DC
Range
PeakTech 1006: 2 Hz...6 MHz
PeakTech
1013:
2
Hz...13
MHz
Ranges of frequency
2 - 20 Hz
20
-
200
Hz
200
Hz
-
2
kHz
2 kHz - 20 kHz
20
kHz
-
200
kHz
200
kHz
-
2
MHz
2 MHz - 6 MHz (PeakTech 1006)
2 MHz - 13 MHz (PeakTech 1013)
Attenuation
0 dB, 20 dB, 40 dB; variable 0 to 20 dB
Resolution
4
digits
Impedance
50
Ω
± 2 %
Amplitude
-10 Vp to +10 Vp into open circuit;
-5 Vp to +5 Vp into 50
Ω
Frequendy Accuracy
± 0.01 % up to 100 kHz