PCAN-Diag 2 – Oscilloscope Function
69
A CSV file contains data in lines in text format. As separator the
semicolon (;) is used. For further use the file can be taken into an
arbitrary spread sheet, for example.
Row
Contents
Structure
1
Device name and firmware version
string
2 Transceiver
type
string
3 - 4
Signal source measuring channels 1 and 2
string
5 Number
of
samples
name;count
6 - 7
Voltage scale measuring channels 1 and 2
name;value
8 - 9
Voltage offset measuring channels 1 and 2
name;value
10
Time scale samples values [s]
name;value
11
Time offset samples values [s]
name;value
12
Column name for the following sample values
name;name;name
13+ Numbered
sample
values
number;value;value
Calculations for a sample value (in brackets: row):
Time:
Time(13+) * Timebase(10) + Time Offset(11)
Voltage, for measuring channels 1 and 2 each:
Channel(13+) * Scale Channel(6/7) + Offset Channel(8/9)