Page 11
Technical Specification
NG3000A
Environment
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
2000 m
Maximum Relative Humidity
80% Non-Condensing
Minimum Storage Temperature*
10°C (50°F)
Maximum Storage Temperature*
25°C (75°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature
for a minimum of 3 hours before operation.
Generator Outlets
Maximum Gas Output Pressure
5.5 bar (80 psi)
Maximum Outlet Flow Rate
3000 cc/min (0.10 cfm)
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
1 x ¼” BSPP
Start-Up Time For Purity
8 hours
Electrical Requirements
Voltage
230v or 110v
Frequency
50/60 Hz
Current
3.6 (8) A
Pollution Degree
2
Insulation Category
II
General
Dimensions cm (inches) WxDxH
43 (16.9) x 41 (16.1) x 1235 (48.6)
Generator Weight Kg (lbs)
88 (194)
Shipping Weight Kg (lbs)
128 (282)