Page 10
Technical Specification
NG3000
Environment
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
25°C (75°F)
Maximum Altitude
2000 m
Maximum Relative Humidity
80% Non-Condensing
Minimum Storage Temperature*
10°C (50°F)
Maximum Storage Temperature*
25°C (75°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature
for a minimum of 3 hours before operation.
Inlet Conditions
Minimum Air Pressure
8.2 bar (120 psi)
Maximum Air Pressure
8.8 bar (130 psi)
Generator Outlets
Maximum Gas Output Pressure
5.5 bar (80 psi)
Maximum Outlet Flow Rate
3000 cc/min (0.10 cfm)
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
1 x ¼” BSPP
Start-Up Time For Purity
8 hours
Electrical Requirements
Voltage
230v or 110v
Frequency
50/60 Hz
Current
0.5 (1.2) A
Pollution Degree
2
Insulation Category
II
General
Dimensions cm (inches) WxDxH
43 (16.9) x 41 (16.1) x 1235 (48.6)
Generator Weight Kg (lbs)
77 (170)
Shipping Weight Kg (lbs)
107 (236)