Rear Connections
The Nitrogen Generator should be connected to a clean, dry source of compressed
air. A minimum pressure of 120 psi is required. Any doubts as to the suitability of the
compressed air supply should be referred to the factory for advice.
The generator has two ¼” BSPT bulkhead connections to the left side of the unit.
The upper port is the Compressed Air inlet and the lower is the Nitrogen outlet. A
¹
/8
” Swagelok fitting is supplied for connection to the application. There is no drain
on this machine. Any moisture liberated by the filter separator is discharged through
the high capacity exhaust system, where sudden reduction in pressure causes instant
evaporation. The water vapour is safely removed from the unit by the ventilation
system.