Commissioning
With the Generator installed (as described earlier), connect compressed air to the unit
and open the air supply to the Unit. Disconnect the Outlet connections to allow the
generator to vent to atmosphere until the unit is stabilised. The Generator has been pre-
set in the factory to give the specified output flow-rate and pressure in line with known
requirements. Once the Membranes reach the design pressures the Generator will
stabilise and produce Nitrogen. Maximum purity will be achieved after approximately 1
hour. The generator can then be re-connected to the application.
The design of the generator is such that it can deliver up to the supplies specified in
the Pressure & Flow Settings table below. Should the demand for gasses be less than
the rated output flow at any time, the demand will be determined by the consuming
equipment. In circumstances of no demand, the Generator remains operational and will
provide Nitrogen immediately as demand resumes.