Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Performance of the generator is affected by ambient conditions. Note should also be
taken to the proximity of Air Conditioning outlets. These can sometimes give rise to
“pockets” of air with high relative humidity. Operation of the unit within such a pocket
could adversely affect its performance. Consideration should also be given to the
air flow around the unit. It is recommended that an air gap of 75mm (3”) should be
maintained down both sides and at the rear of the unit. Please refer to the drawing
below for the general dimensions of the unit.
Minimum Operating Ambient Temperature:
5 °C (35 °F)
Maximum Operating Ambient Temperature:
41 °C (95 °F)