Page 16
Air Connection
Infinity 1045 Generator should be connected to a clean, dry, OIL - FREE source
of compressed air. A minimum inlet pressure of 116 psig (8 barg) is required. Any
doubts as to the suitability of your compressed air supply should be referred to Peak
Scientific or any of their authorised partners.
The generator has a Breathing Air Filter with ¼” BSPT connection to the left side
of the unit. The Compressed Air supply should be connected here. This filter will
drain moisture and is equipped with an automatic drain. The drain should be led to a
convenient place. There are 2-off ¼” BSPT Female bulkhead connections to the right
of the unit. Fittings are provided to connect these ports to the gas connections of the
Mass Spectrometer.
LC/MS with 2-off Connections (e.g. LC-TQ)
Connect the 2-off gas ports at the right side of the generator to their respective ports
on the Mass Spectrometer using the 2-off ¼” BSPT x ¼” Swagelok fittings. Use ¼”
tubing throughout.