![PCE Instruments PCE-IT100 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/pce-instruments/pce-it100/pce-it100_user-manual_1516140008.webp)
© PCE Instruments
6
Insulation measurement at 1000 V
(0 … ±10 %)
Measurement range
1.
000 … 4.000 MΩ
4.
001 … 40.00 MΩ
40.
01 … 400.0 MΩ
400.
1 … 4000 MΩ
Resolution
0.001
MΩ
0.01
MΩ
0.1
MΩ
1
MΩ
Accuracy
±(3 % + 10 digits)
±(2 % + 10 digits)
±(2 % + 5 digits)
±(4 % + 5 digits)
Test current
1 mA at k
Ω
Short circuit current
≤1 mA
General specifications
Continuity test
Audible signal at resistance
<35 Ω
Battery indication
„
“ is displayed when battery voltage is no longer sufficient
Excess of measurement range
“OL“ is displayed
Sampling rate
2.5 measurements/s (0.4 Hz)
Zero point
Manual adjustment possible
Power supply
6 x 1.5 V AAA
Fuse
10 A / 600 V (5 x 20 mm)
Operating conditions
0 ... +40 °C, <80 % RH
32 … 104 °F
Storage conditions
-10 ... +60 °C, <70 % RH
14 … 140 °F
Operating altitude
2000 m
Dimensions
200 x 92 x 50 mm
Weight
700 g with batteries
Standards
IEC10101, CAT III 1000 V, pollution degree 2