Continuity test/resistance measurement 

To make a continuity test or resistance measurement, select 

“400Ω“ with the rotary switch. You 

can now connect the test leads to your sample and measure the resistance. The measurement 
range is set automatically. The continuity test takes place at the same time. 

Note: 

If the sample carries a voltage of at least 30 V, the meter will display 

“>30 V“ and “ “, a 

beep sound will be audible and the meter will not make a measurement. Therefore, it is important 
to remove any external voltages before each measurement. 
 

Voltage measurement AC/DC 

To measure alternating current, select 

“750V“ with the rotary switch. Now you can connect the 

test  leads  to  your  sample.  The  present  voltage  will  directly  be  shown  in  the  upper  part  of  the 
display. The current battery voltage will be shown in the lower part of the display. 
To measure direct current, turn the rotary switch to the 

”1000V“ position. Then connect the test 

lead to the sample. The reading will be shown directly in the upper part of the display. 
The measurement range is determined automatically for both functions. 
 

10  Automatic Power Off function 

The meter turns off automatically after 10 minutes of inactivity. This function cannot be disabled. 
To turn on the meter again after auto power off, turn the rotary switch back to 

“OFF“ and then turn 

it o the desired position. 
 

11  Keys 

The meter has four keys which have the following functions: 

11.1 

Hold reading 

Press the HOLD key to freeze the reading. 

“HOLD“ will be displayed. Press the HOLD key again 

to continue with the measurement. 

“HOLD” will now disappear. 

11.2 

MAX/MIN 

Press and hold the MAX/MIN key to start the MAX/MIN function. 

“MAX“ will appear in the display 

first. This function will display the maximum reading. Press the key again to view the lowest value. 
“MIN” will be displayed. Press and hold this key to return to normal measuring mode. 

Note: 

This  function  is  only  available  with  the  measuring  functions 

”400Ω“,  ”1000VDC“  and 

”750VAC“. 

11.3 

Freehand measurement (LOCK key)  

Press the LOCK key. A lock icon will appear in the display. Now press the TEST key. An audible 
signal indicates that voltage is now present in the measuring tips. This voltage will be shown as 
a reading in the lower part of the display. The currently measured resistance value is shown in 
the upper part of the display. The high voltage is discharged as soon as the TEST key is pressed 
again.  This  procedure  can  be  followed  by  looking  at  the  lower  part  of  the  display.  When  the 
discharge is completed, the beep sound will be muted and the measuring tips can be removed 
safely. 
 

 

Содержание PCE-IT100

Страница 1: ...sulation Tester User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...o point 8 7 Insulation measurement 8 7 1 Measuring AC motors 8 7 2 Measuring DC motors 8 7 3 Measuring insulated cables 8 8 Continuity test resistance measurement 9 9 Voltage measurement AC DC 9 10 Au...

Страница 3: ...es from PCE Instruments or equivalent Before each use inspect the case for visible damage If any damage is visible do not use the device Do not use the instrument in explosive atmospheres The measurem...

Страница 4: ...ons on the meter which mean the following This icon can be found next to another symbol or a connection and is a reference to the user manual This icon indicates that high voltage could be present Sho...

Страница 5: ...ts in commercial buildings Motors switches sub distributors in a three phase circuit light circuits in commercial buildings supply cables for industrial plants electrical appliances or connections nea...

Страница 6: ...0 00 400 0 Resolution 0 01 0 1 Accuracy 1 2 3 digits Overvoltage protection 250 V RMS Measuring voltage max 5 8 V Continuity test Icon Resolution 0 01 Audible signal 35 Short circuit current 200 mA Ov...

Страница 7: ...ment at 250 V 0 10 Measurement range 0 250 4 000 M 4 001 40 00 M 40 01 400 0 M 400 1 4000 M Resolution 0 001 M 0 01 M 0 1 M 1 M Accuracy 2 10 digits 2 10 digits 5 5 digits 4 5 digits Test current 1 mA...

Страница 8: ...st Audible signal at resistance 35 Battery indication is displayed when battery voltage is no longer sufficient Excess of measurement range OL is displayed Sampling rate 2 5 measurements s 0 4 Hz Zero...

Страница 9: ...truments 7 4 Device description 1 LCD 2 MAX MIN HOLD key 3 LOCK key 4 Zero point and backlight key 5 TEST key 6 Rotary function switch 7 V input jack 8 Ground input jack 9 Strap holder 10 Battery comp...

Страница 10: ...sent in the meter The reading will be shown in the upper part of the display The test voltage is shown in the lower part of the display The measurement range is selected by the meter automatically Not...

Страница 11: ...es of inactivity This function cannot be disabled To turn on the meter again after auto power off turn the rotary switch back to OFF and then turn it o the desired position 11 Keys The meter has four...

Страница 12: ...ttery compartment at the rear side the cover of which is affixed with four screws Remove all batteries You can now replace the fuse Only use an FF 500 mA 1000 V fuse 15 Warranty You can read our warra...

Страница 13: ...583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info pce instruments co uk info pce italia it www pce instruments com deutsch www pce instrumen...

Отзывы: