![PCE Instruments PCE-IT100 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/pce-instruments/pce-it100/pce-it100_user-manual_1516140007.webp)
© PCE Instruments
5
Insulation measurement at
125 V (0 … ±10 %)
Measurement range
0.
125 … 4.000 MΩ
4.
001 … 40.00 MΩ
40.
01 … 400.0 MΩ
400.
1 … 4000 MΩ
Resolution
0.001
MΩ
0.01
MΩ
0.1
MΩ
1
MΩ
Accuracy
±(2 % + 10 digits)
±(2 % + 10 digits)
±(4 % + 5 digits)
±(5 % + 5 digits)
Test current
1 mA at 125 k
Ω
Short circuit current
≤1 mA
Insulation measurement at 250
V (0 … ±10 %)
Measurement range
0.
250 … 4.000 MΩ
4.
001 … 40.00 MΩ
40.
01 … 400.0 MΩ
400.
1 … 4000 MΩ
Resolution
0.001
MΩ
0.01
MΩ
0.1
MΩ
1
MΩ
Accuracy
±(2 % + 10 digits)
±(2 % + 10 digits)
±(5 % + 5 digits)
±(4 % + 5 digits)
Test current
1 mA at 125 k
Ω
Short circuit current
≤1 mA
Insulation measurement at 500 V
(0 … ±10 %)
Measurement range
0.
500 … 4.000 MΩ
4.
001 … 40.00 MΩ
40.
01 … 400.0 MΩ
400.
1 … 4000 MΩ
Resolution
0.001
MΩ
0.01
MΩ
0.1
MΩ
1
MΩ
Accuracy
± (2 % + 10 digits)
± (2 % + 10 digits)
± (2 % + 5 digits)
± (4 % + 5 digits)
Test current
1 mA at 500 k
Ω
Short circuit current
≤1 mA