This section provides some basic steps to identify possible causes of trouble, together with troubleshooting
solutions. The FT-20 is a sealed unit. If there are any problems with the PowerPak which go beyond the scope of
these instructions, contact Fatigue Technology Inc. or an FTI Representative who can provide you with the nearest
authorized service center.
WARNING:
To prevent injuries, any repair work or troubleshooting must be done by qualified personnel familiar with
this equipment. Use proper equipment and tools when troubleshooting.
IMPORTANT: Refer to the parts lists (Section 6.0) when using the Troubleshooting Guide.
PROBLEM
CAUSE
SOLUTION
5.1 PUMP WILL NOT START OR WILL NOT STOP REGARDLESS OF TRIGGER POSITION
NOTE:
This is common for new pumps or different puller unit/pump combinations.
(a) Trigger response valve is not
properly adjusted.
(a) While squeezing the trigger of the puller unit,
loosen the brass nut on the trigger response
valve. Adjust the valve pilot in or out until the
desired sensitivity is obtained. Retighten the
brass nut.
(b) Air logic valve is stuck in
position.
(b) The air logic valve is located on top of the pump
just under the shroud. Remove the shroud and
disassemble the valve. Remove the valve spool
and inspect the O-rings and spool body. If the
valve is a Versa
brand valve (gold body),
replace the O-rings as required and remove any
contaminates. If the valve is an I.S.I. brand valve
(black body), the spool will most likely need to
be replaced. Clean any contaminates from the
valve body. Refer to Section 6.0 for valve part
numbers.