NMS044 Reference Manual Rev E
Default NMS044 Setup File
4-3
FIGURE 4-3
NMS044 Settings: Control
TAKE NOTE
Measurement
history is only available with
the 831C-ELA or 831C-MRS
option purchased and
installed on the Model 831C
SLM. For more information,
see “Contact Larson Davis” on
page i-2.
It is recommended to set Continuous as the Run Mode for the NMS044.
This mode will begin a run every time the meter is powered on and will
continue to run until a manual Stop or the meter is powered off. If the
system loses power and then is recharged sufficiently, the system will
power on and a measurement will start on its own. The Daily Auto Store
is set to store once a day, so you will have 24 hour measurement files.
The PRM2103 preamplifier can do a daily automatic calibration check
to ensure that the measurements are accurate. A recommended time
for the calibration check is 2:30:00.
FIGURE 4-4
NMS044 Settings: Time History
TAKE NOTE
Time History is only
available with the 831C-LOG
or 831C-MRS option
purchased and installed on
the Model 831C SLM. For more
information, see “Contact
Larson Davis” on page i-2.
Time History at one second period will create one second records in
each measurement. For each record, the selected metrics will be
logged. The NMS044 setup enables the default metrics.