Carbon Dioxide ISE Probe

Replacement Items

4

012-15719A

Connect the electrode to the sensor or measuring 

device. Remove the cover from the electrode body.

2. Prepare two standard solutions that differ in 

concentration by a factor of ten and bracket the 

expected sample concentration range. For example, 

if your expected sample concentration is 50 mg/L 

Carbon Dioxide, you should use a 10 mg/L low 

Standard Carbon Dioxide Solution and a 100 mg/L 

high Standard Solution. Standards can be prepared 

in any concentration unit to suit the particular analy

-

sis requirement. 

NOTE: All standard solutions should be at the same 

temperature as the sample.

Measurement:

If using a measuring device with direct concentration 

reading capability, see the software on-line help or the 

measuring device’s instruction manual for specific 

information:

1. Place 100 mL of the low standard into a 150 mL 

beaker. Add 1 mL of ISA. Stir thoroughly. Check the 

pH value of the solution.

2. Rinse electrode with Dl water, blot dry and place in 

the beaker. Wait for a stable reading, and then 

adjust the software or the measuring device to 

display the value of the standard. Refer to the soft

-

ware on-line help or the measuring device’s 

instruction manual for the adjustment procedure.

3. Measure 100 mL of the high standard into a second 

150 mL beaker. Add 1 mL of ISA and stir. Check the 

pH value of the solution.

4. Rinse electrode with Dl water, blot dry and place in 

the second beaker. Wait for a stable reading, and 

then adjust the software measuring device to display 

the value of the second standard.

5. Measure 100 mL of the sample into a 150 mL 

beaker. Add 1 mL of ISA. Stir thoroughly. Check the 

pH value of the standard.

6. Rinse electrode with Dl water, blot dry and place in 

the sample beaker. Wait for a stable reading and the 

sample concentration will be displayed on the mea

-

suring device.

7. Determine the sample concentration using the data 

analysis tools in the software.

Electrode Storage

Short Term (over night):

Between measurements, keep the electrode tip 

immersed in a 10 ppm standard without ISA. For 

low-level measurements, keep the electrode in pH 7.00 

buffer between measurements. For overnight storage, 

the electrode tip should be immersed in pH 7.00 buffer 

solution. 

Long Term

For storage over one week, or if the electrode is stored 

indefinitely, disassemble completely and rinse the Inter

-

nal Glass Electrode and Electrode Body with DI water. 

Dry all the parts. Do not re-insert the Internal Glass Elec

-

trode into the Electrode Body. Instead, cover the tip of 

the Internal Glass Electrode with the original boot. Store 

all the parts separately in the original box. When reas

-

sembling, follow the procedure in the “Electrode Assem

-

bly” and “Checking Electrode Operation” seconds. 

Life Span

The PS-3517 Carbon Dioxide ISE Probe is expected to 

last more than three years before it needs to be 

replaced.

Replacement Items

PS-3517 Carbon Dioxide ISE Probe

SC-3528 Carbon Dioxide ISE Solutions: 60-mL 

Standard Solution, 30-mL Reference Fill Solution, 

30-mL ISA Solution

Technical Support

For assistance with any PASCO product, contact 

PASCO at:

Limited Warranty

For a description of the product warranty, see the PASCO catalog.

Address:

PASCO scientific

10101 Foothills Blvd.

Roseville, CA 95747-7100

Phone:

916-462-8384 (worldwide)

800-772-8700 (U.S)

Email:

[email protected]

Web:

www.pasco.com/support

Содержание PS-3517

Страница 1: ...d area Boot 1 Cover DO NOT let the BNC connector at the top of the unit get wet Included Equipment Item Included Equipment Item Carbon Dioxide ISE Electrode Body 2 1 CO2 Standard Solution 1000 ppm 4 Carbon Dioxide ISE Internal Electrode 2 CO2 Reference Fill Solution 5 BNC Male to Female Cable 3 CO2 Ionic Strength Adjuster ISA 6 ...

Страница 2: ...e tip of the internal glass electrode Keep the boot for later use Note Soaking the internal glass electrode in internal filling solution or pH 4 01 buffer for at least 1 hour before proceeding with the following steps will ensure fast response 2 Remove an electrode body from the box Using the solution provided fill the electrode body with about 3 5 mL of Reference Fill Solution 3 Rinse the interna...

Страница 3: ...ctrode response follow the steps below 1 Check the slope of the internal glass electrode Remove the internal glass electrode from the electrode body Connect the internal glass electrode to the Wireless pH Sensor or to a pH mV meter Put the internal glass electrode in pH 7 00 buffer The annular ceramic junction should be immersed in the solution Record the reading in mV when it is stable Remove the...

Страница 4: ...100 mL of the sample into a 150 mL beaker Add 1 mL of ISA Stir thoroughly Check the pH value of the standard 6 Rinse electrode with Dl water blot dry and place in the sample beaker Wait for a stable reading and the sample concentration will be displayed on the mea suring device 7 Determine the sample concentration using the data analysis tools in the software Electrode Storage Short Term over nigh...

Страница 5: ...ay be trademarks or service marks of and are used to identify products or services of their respective owners For more information visit www pasco com legal Product End of Life Disposal Instructions This electronic product is subject to disposal and recycling regulations that vary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws...

Отзывы: