Carbon Dioxide ISE Probe

Introduction

2

012-15719A

Introduction

This Carbon Dioxide Ion Selective Electrode (ISE) Probe 

is a gas-sensing electrode that allows fast, simple, eco

-

nomical, and accurate measurement of dissolved CO

2

 in 

aqueous solutions. 

Technology: Replaceable Electrode Body with PVC 

Membrane 

Overview

Required Equipment

PASCO Wireless pH/ISE/ORP Sensor (PS-3204)

or 

PASPORT High Precision pH/Temperature with 

ISE/ORP Amplifier (PS-2107) with a PASCO 

interface (see www.pasco.com for information)

or

Ion meter or pH/mV meter

PASCO data collection software (see 

www.pasco.com)

Wash bottle filled with distilled or de-ionized water 

pH Buffer Solution 4.01* 

pH Buffer Solution 7.00* 

Lab wipes 

Magnetic Stirrer with Stir Bar

Clean beakers**

(*SC-2321 pH Buffer Capsule Kit)

(**SE-7287 100 mL Beakers (12 pack) or SE-7288 1000 

mL Beakers (6 pack)

Included Required Solutions*

Reference Fill Solution

1000 ppm Carbon Dioxide Standard Solution

Ionic Strength Adjuster (ISE)

(*For replacement items, see the Technical Support sec

-

tion at the end of this document.)

Electrode assembly 

Important. Please read these instructions before assem

-

bling the Carbon Dioxide ISE Probe

1. Carefully remove the internal glass electrode from 

the box. Remove the boot, which covers the tip of 

the internal glass electrode. Keep the boot for later 

use.

Note: Soaking the internal glass electrode in internal 

filling solution or pH 4.01 buffer for at least 1 hour 

before proceeding with the following steps will 

ensure fast response. 

2. Remove an electrode body from the box. Using the 

solution provided, fill the electrode body with about 

3.5 mL of Reference Fill Solution.  

3. Rinse the internal glass electrode with DI water, blot 

dry. Place it into the electrode body, which has been 

filled with the filling solution.

IT IS VERY IMPORTANT TO MAKE SURE THAT THE 

KEY PIN ON THE INTERNAL GLASS ELECTRODE 

CAP MUST FIT INTO THE NOTCH AT THE TOP 

OPENING OF THE ELECTRODE BODY SO THAT THE 

INTERNAL GLASS ELECTRODE WILL NOT SPIN 

DURING FURTHER ASSEMBLY. THE INTERNAL 

GLASS ELECTRODE SPINNING DURING ASSEMBLY 

WILL DAMAGE THE MEMBRANE AT THE BOTTOM 

OF THE ELECTRODE BODY, WHICH IS NOT 

COVERED BY THE WARRANTY. 

4. Install the threaded top cap onto the threaded sec

-

tion at the top of the electrode body. Tighten the cap 

until snug. Do not over tighten. 

Size and Material

Body (Epoxy): 12 mm OD x 155 

mm Length

Cap (ABS): 16 mm OD x 47 mm 

Length

Range

4.4 to 400 ppm as CO

2

Reproducibility

±2% of reading

Slope

52 to 60 mV at 25°C, typical

Minimum Sample

3 mL in a 30 mL beaker

pH Range

Samples and standards must be 

adjusted to below pH 4.00

Operating Temperature 0 to 50°C

Electrode resistance

less than 1,000 megaohms

Interferences

volatile weak acid (gaseous)

Содержание PS-3517

Страница 1: ...d area Boot 1 Cover DO NOT let the BNC connector at the top of the unit get wet Included Equipment Item Included Equipment Item Carbon Dioxide ISE Electrode Body 2 1 CO2 Standard Solution 1000 ppm 4 Carbon Dioxide ISE Internal Electrode 2 CO2 Reference Fill Solution 5 BNC Male to Female Cable 3 CO2 Ionic Strength Adjuster ISA 6 ...

Страница 2: ...e tip of the internal glass electrode Keep the boot for later use Note Soaking the internal glass electrode in internal filling solution or pH 4 01 buffer for at least 1 hour before proceeding with the following steps will ensure fast response 2 Remove an electrode body from the box Using the solution provided fill the electrode body with about 3 5 mL of Reference Fill Solution 3 Rinse the interna...

Страница 3: ...ctrode response follow the steps below 1 Check the slope of the internal glass electrode Remove the internal glass electrode from the electrode body Connect the internal glass electrode to the Wireless pH Sensor or to a pH mV meter Put the internal glass electrode in pH 7 00 buffer The annular ceramic junction should be immersed in the solution Record the reading in mV when it is stable Remove the...

Страница 4: ...100 mL of the sample into a 150 mL beaker Add 1 mL of ISA Stir thoroughly Check the pH value of the standard 6 Rinse electrode with Dl water blot dry and place in the sample beaker Wait for a stable reading and the sample concentration will be displayed on the mea suring device 7 Determine the sample concentration using the data analysis tools in the software Electrode Storage Short Term over nigh...

Страница 5: ...ay be trademarks or service marks of and are used to identify products or services of their respective owners For more information visit www pasco com legal Product End of Life Disposal Instructions This electronic product is subject to disposal and recycling regulations that vary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws...

Отзывы: