background image

®

W e a t h e r / A n e m o m e t e r   S e n s o r

  P S - 2 1 7 4

4

where 

 is the saturation 

water vapor pressure, RH is Relative Humidity (as a percentage), 
and 

T

 is Temperature in °C. 

Absolute Humidity

Absolute Humidity is a measure of the 

mass of water vapor present in a given volume. It is reported in 
units of g/m

3

. The sensor calculates Absolute Humidity with the 

formula:

Where RH is Relative Humidity (as a percentage), and 

T

 is Tem-

perature in °C.

Wind Chill

Wind Chill indicates a combined effect of temper-

ature and wind speed. It is reported as an apparent temperature on 
the Celsius and Fahrenheit scales. The sensor calculates Wind 
Chill (in °F) using the formula:

where 

T

 is air temperature (in °F) and 

V

 is wind speed (in miles 

per hour).

Wind Chill is defined for temperatures at or below 50° F and 
wind speed above 3 miles per hour; outside of those limits, the 
sensor reports a value equal to Air Temperature.

Humidex

Humidex indicates a combined effect of temperature 

and humidity. It is reported as an apparent temperature on the 
Celsius and Fahrenheit scales. The sensor calculates Humidex (in 
°F) using the formula:

where 

T

 it temperature in °C, and 

e

 is the water vapor pressure in 

mbar.

Sea-level Corrected Pressure

A typical weather station does not report the actual measured 
barometric pressure. Rather, it reports a “sea-level corrected” 
value. Use this formula to calculate the sea-level corrected pres-
sure (in mbar):

[Barometric Pressure (mbar)]+1013*(1-exp(

-h

/7000))

where 

h

 is the sensor’s elevation in m. To have this calculation 

done automatically, enter it in the DataStudio calculator or GLX 
Calculator screen.

Specifications

Compatible Interfaces

The Weather/Anemometer Sensor is compatible with all wired 
PASPORT interfaces and loggers including USB Link (PS-2100), 
Xplorer (PS-2000), PowerLink (PS-2001), and Xplorer GLX 
(PS-2002).

For wireless use, the Weather/Anemometer Sensor is compatible 
with AirLink SI model PS-2005A. It is not compatible with 
model PS-2005.

Technical Support

For assistance with any PASCO product, contact PASCO at:

Limited Warranty

For a description of the product warranty, see the 

PASCO catalog.

Copyright

The PASCO scientific 012-09911D 

Weather/Anemometer 

Sensor Instruction Sheet

 is copyrighted with all rights reserved. Permis-

sion is granted to non-profit educational institutions for reproduction of any 
part of this manual, providing the reproductions are used only in their labo-
ratories and classrooms, and are not sold for profit. Reproduction under 
any other circumstances, without the written consent of PASCO scientific, 
is prohibited.

Trademarks

PASCO, PASCO scientific, DataStudio, ScienceWorkshop, 

Xplorer and Xplorer GLX are trademarks or registered trademarks of 
PASCO scientific, in the United States and/or in other countries. All other 
brands, products, or service names are or may be trademarks or service 
marks of, and are used to identify, products or services of, their respective 
owners. For more information visit www.pasco.com/legal.

SatVP

6.11

7.75

T

237.7

T

+

10

=

Absolute Humidity

13.24 RH

T

273.15

+

----------------------------

17.42

T

T

239.7

+

-----------------------

exp

=

Wind Chill (°F)

35.74 0.6215

T

35.75

V

0.16

0.4275

T

0.16

+

+

=

Humidex

T

5
9

---

e

10

+

=

Measurement

Range

Resolution

Accuracy

(% of reading)

Wind Speed

0.5 to 29 m/s

0.1 m/s

±(3% of reading 

+ 0.2 m/s)

Temperature

-20 to 55 °C

0.1 °C

±0.5 °C

Relative 
Humidity

0 to 100%

1%

±2%

Barometric 
Pressure

150 to 1150 hPa

0.03 hPa

1 hPa

Address: PASCO scientific

10101 Foothills Blvd.
Roseville, CA 95747-7100

Phone:

916-786-3800 (worldwide)
800-772-8700 (U.S.)

Fax:

(916) 786-7565

Web:

www.pasco.com

Email:

[email protected]